Operating Manual
ALARMS AND PROBLEM CORRECTION
Software 2.6
B|BRAUN
Chapter 8
Page 15 of 27
SHARING EXPERTISE
2.6_CA-EN_03
Code
Warning Text
Reason for Warning
Corrective Action
W 12
Plasma reinfusion
completed !
For Blood Reinfusion Stop
Blood Pump
(do not press 'OK')
or
for further Plasma
Reinfusion
Press ‘OK’ to proceed !
Plasma reinfusion
completed,
information
concerning the
preparation for blood
reinfusion
Follow the instructions on the monitor to
change to the blood reinfusion or press
the
key to continue plasma
reinfusion.
M
W 14
Compulsory rinsing
completed. Set new value
to continue rinsing.
The minimum rinsing
volume of 2400 ml has
been reached.
Confirm the warning with the
key.
Change to the therapy mode when you
consider the rinsing volume to be
sufficient.
Increase the rinsing volume (see chapter
4) and therefore extend the rinsing phase,
if required (e.g. when replacing a filter
during the rinsing phase).
M
W 15
Connect buffer – seal and
clamp opened ?
Confirmation before
the start of the
therapy.
Check the positions given on the monitor
and confirm with the
key to
continue.
M
W 16
Press 'OK' to return to
menu selection !
Information for
quitting the screen
when adjusting the
parameters
Press the
key to return from
<
Parameter Setting
> to the menu bar.
T
W 17
Blood reinfusion
completed !
Information that blood
reinfusion is
completed.
Remove the venous line from the patient
and terminate the treatment.
Increase the blood reinfusion volume (see
chapter 6) and continue reinfusion if you
consider it necessary.
T
W 18
Break seals and open all
clamps !
Confirmation at the
start of priming and
rinsing
Follow the instructions on the monitor
and confirm with the
key to
continue.
M
W 19
Press 'OK' to exclude BLD
alarms !
Is offered as an option
after three BLD alarms
Press the
key to override the BLD
alarm.
M
W 20
BLD alarms excluded !
Information when the
BLD alarm has been
overridden by
accepting the W19
option.
T
W 21
1) Connect art. line to
saline solution bag
2) Connect reinfusion line
to venous chamber
Confirmation before
the blood reinfusion.
Check the positions given on the monitor
and confirm with the
key to
continue.
M
OK
OK
OK
OK
OK
OK
OK