Operating Manual
THERAPY
Software 2.6
B|BRAUN
Chapter 5
Page 5 of 14
SHARING EXPERTISE
2.6_CA-EN_03
CAUTION
Risk to patient due to haemolysis because of a
high shear stress.
Gradually increase first the blood flow rate to
reach the desired target value after 5 minutes.
Subsequently, gradually increase the plasma
flow rate to achieve the appropriate value
after another 10 minutes.
CAUTION
Risk to patient due to insufficient dialysis after
restart because of spontaneous ultrafiltration
occuring in the therapy stand-by mode.
Place a clamp on the dialysate drainage line
behind the dialyzer.