Operating Manual
PRIMING AND RINSING
Software-Version 2.6
B|BRAUN
Chapter 4
Page 6 of 19
SHARING EXPERTISE
2.6_CA-EN_03
Step 4
The plasma/buffer pump starts and the precipitate
filter is filled. This step is completed when the level
monitoring of the precipitate filter air chamber (PCLD)
detects fluid and the balance test 1 has been
completed.
Step 5
Filling the heparin adsorber air chamber (HCLD)
Step 6
Leakage test of the heparin adsorber clamp
Step 7
The heparin adsorber clamp (HAK) opens. The level
detection in the heparin adsorber air chamber and the
venting of the connection line to the heparin adsorber
are performed. This step includes the filling of the
dialyzer on the plasma side.
Step 8
The Warning window prompts with<
W04: Turn
dialyzer (blue side down) !
>.
Turn the dialyzer by 180
o
, with the blue side pointing
downward.
Press the
key to continue.
Step 9
The dialysate side filling of the dialyzer is performed
during this step.
The balance test 2, the DAD test, the heating test
,
the
venous pressure test as well as the reinfusion pump
test are performed in this step.
Step 10
The setting of the level of the venous air chamber is
performed.
OK
Warning
W04: Turn dialyzer (blue side down) !
Press ‘OK’ to proceed !