Operating Manual
THERAPY
Software 2.6
B|BRAUN
Chapter 5
Page 6 of 14
SHARING EXPERTISE
2.6_CA-EN_03
5.2 Terminating the Therapy
When the treated plasma volume is achieved, the
machine switches to the stand-by mode. The blood
circuit continues to circulate with the most recent
blood flow rate selected.
The cursor automatically points to the <
End of
Therapy
> command in the menu bar. Confirm
with the
key.
Confirm the message <
W35: Activate
reinfusion? Press 'OK' to proceed!>
in the
Warning window by pressing the
key to
change to the reinfusion phase.
CAUTION
Risk to patient due to blood and/or plasma loss and
subsequent blood pressure drop in case of a
premature termination of the therapy without
reinfusion of the plasma and/or blood volume.
Volume substitution, application of an albumin
solution as prescribed by the handling
physician.
Request the patient to drink more liquids than
usual.
Status 2005-10-26
OK
Warning
W35: Activate reinfusion ?
Press ‘OK’ to proceed !