Operating Manual
PRIMING AND RINSING
Software-Version 2.6
B|BRAUN
Chapter 4
Page 7 of 19
SHARING EXPERTISE
2.6_CA-EN_03
Step 11
This step is completed when the minimum rinsing
volume of 2400 ml is reached. The following message
is displayed in the Warning window: <
W14: Rinsing
completed. Set new value to continue rinsing
Continue with 'OK!
>
Press the
key to confirm the reaching of
the minimum rinsing volume.
If the minimum rinsing volume is sufficient, you
can now start with the therapy.
Step 12
Optional rinsing
This step allows rinsing of the system beyond the
minimum rinsing volume.
If you wish to increase the rinsing volume:
Select the <
Parameter Setting
> command in the
menu bar.
Select the <
Rinsing volume
> parameter and
change this parameter. The rinsing volume can be
set to a value of up to 20 l.
Then select the <
Start Priming
> command in the
menu bar. When the rinsing volume has been
reached, all pumps stop automatically.
For more details on increasing the rinsing volume, see
also the Chapters 4.2.1 and 4.2.2.
When increasing the rinsing volume over 2400 ml,
make sure that sufficient saline solution is available.
If required, change the bags on the load cell and on
the IV-pole.
OK
W14: Rinsing completed. Set new value to continue
rinsing
Press ‘OK’ to proceed !
Warning
W14: Rinsing completed. Set new value to continue
rinsing