Operating Manual
REINFUSION
Software 2.6
B|BRAUN
Chapter 6
Page 2 of 14
SHARING EXPERTISE
2.6_CA-EN_03
6.1 PLASMA REINFUSION
After terminating the therapy as described in chapter
5.2, the screen display changes to the Reinfusion
screen.
Display Area of the Reinfusion Screen
Blood flow in ml/min
Blood reinfusion time in min
Blood reinfusion volume in ml
Reinfusion flow in ml/min
Reinfusion time in min
Reinfusion volume in ml
Therapy time [Actual/Rest] in hh:mm
Plasma volume [Actual/Rest] in ml
Arterial pressure in mmHg
Venous pressure in mmHg
After the change to the reinfusion phase, the blood
flow is not stopped but set automatically to
40 ml/min.
The default setting of the plasma reinfusion volume is
400 ml.