2-3-2. Von Voltage Settings
2-3-2-1. Von Voltage Level
Description
The Von Voltage is the threshold voltage at which the
load module will start to sink current.
Operation
Press
Main
>
Configure
[F5] >
Other
[F2] in order,
and set the
Von Voltage
level.
Von Voltage:
0.00 ~
rated voltage
2-3-2-2. Von Voltage Delay
Description
Von Delay is the amount of time the unit will wait before
turning the load on after the Von Voltage threshold has
been latched. This will prevent overshoot current from
affecting the Von Voltage threshold.
Operation
Press
Main
>
Configure
[F5] >
Other
[F2] in order,
and set the
Von Delay
time.
Von Delay:
OFF, 2.0~60ms
Von voltage
Voltage
Time
Input voltage
Current
Von
Input current
Time
Load ON