28
1-4-13. Remote Sense
Description
Remote sense can be used to help compensate for long
cable length. The longer the cable, the higher the
potential resistance and inductance, therefore a short
cable is best. Twisting the cable can help reduce induced
inductance and using the remote sensing terminals
compensates the voltage drop seen across the load
leads, especially leads with higher resistance. This is
useful when used in CV, CR or CP mode.
Steps
1.
Turn the power off from the rear panel or put the unit
into standby mode.
2.
Turn the power off from the power source.
3.
Connect the sense wires to the remote sensing
terminals:
Connect the positive sense (+S) terminal to the high
potential output of the power source.
Connect the negative sense (-S) terminal to the low
potential output of the power source.
DUT
-
+
Programable
Electronic
Load
+
-
+S
-S
Twisted
pair
Power
source