4-1-16. Monitor Signal Output
4-1-16-1. Trigger Signal Output
Description
The trigger output signal is generated every time a switching
operation is performed (i.e., Dynamic mode) or when a Fast
or Normal Sequence is executed and the TRIG OUT
parameter is enabled.
The trigger output signal from TRIG OUT BNC is a 5V pulse
of at least 2us with an impedance of 500Ω. The common
potential is connected to the chassis potential. The signal
threshold level is TTL.
4-1-16-2. Current Monitor Output
Description
The voltage output from the IMON OUT terminal and from
the IMON pin on the J3 connector is used to represent the
current input level.
The V Range used to represent the full scale current range
from the IMON OUT terminal and from the IMON pin on the
J3 connector depends on the I Range settings.
Monitor Connector
I Range
Monitor Output Range
I MON OUT (BNC)
H, L
0 ~ 10V
M
0 ~ 1V
I MON (J3)
H, L
0 ~ 10V
M
0 ~ 1V
0A
Input current
0V
10V or 1V
Rated current
I MON voltage
TRIG OUT: ON
Time
Input current
TRIG OUT
Start of step
or
Switch operation