3-3-4. Running a Normal Sequence function
Description
Turn the road on, the Normal Sequence function is running.
Operation
1.
Press
FUNC
> Normal Sequence
[F2] in order.
2.
Turn normal sequence mode on by setting
N. Seq.
[F1] to
on
.
NSEQ
will appear at the top of the LCD Display
when
N. Seq.
is On.
3.
Turn the load on.
See page 87 (
3-1-2.Turning on the Load with the Selected
Function
) for the load on.
The normal sequence function starts immediately.
The
NSEQ
icon turns orange when the load is turned on.
4.
When a normal sequence function is running, the LCD
displays which sequence number, step number and
number of loop are currently active. It also displays the
elapsed or remaining test time and elapsed/remaining
time of the current step.
Sequences can be paused by pressing
Pause
[F1] and
resumed again by pressing
Continue
[F1].
If no steps have been created “No N.Seq.” will be displayed
on the screen.
“Sequence Complete
” will be displayed at the end of the
sequence.
Display:
Sequence
Running
Continue
NSEQ
RS232
01/Oct/2017
Run N.Seq.
0.000
V
0.0000
A
0.00
W
Step
Seq. No:
01
Loop:
0001
0003
0:00:05
0:00:46
Current step readback
measurements
Step elapsed/
remaining time
Current sequence,
step and loop
number
Elapsed or
remaining time for
the total test time
Note
The combined test time for all sequences will be displayed
as elapsed test time if the elapsed time is >1000 hours, else
the remaining test time will be displayed.
Summary of Contents for LSG-1050H
Page 24: ...14 GRA 414 E EIA standard 132 382 5 477 6 GRA 414 J JIS standard 149 382 5 477 6...
Page 185: ...175 7 6 Operating Area 7 6 1 LSG 175H H Range M Range L Range...
Page 186: ...176 7 6 2 LSG 350H H Range M Range L Range...
Page 187: ...177 7 6 3 LSG 1050H H Range M Range L Range...
Page 188: ...178 7 6 4 LSG 2100H H Range M Range L Range...
Page 204: ......