98
Data Edit
Overview
Each step in a normal sequence contains the following
setting parameters:
Setting
Setting Range
Description
Step
0001 ~ 1000
Selects and displays the current
step in the sequence.
The number of available steps is
dependent on the number of steps
added using the
Insert Point
[F1]
functions.
Value
The current, voltage, power or
resistance setting for the selected
operating mode.
Time
0.05ms -
999h:59m
Sets the step time for the selected
step.
Load
ON, OFF
Turns the load on or off for the
selected step.
RAMP
ON, OFF
When turned on the current
transition is evenly ramped from the
start of the step to the end of the
step.
When turned off the current
transition is stepped.
Ramp: OFF
Time
Input current
Step time
Ramp: ON
Time
Input current
Step time