104
Timing Edit
Overview
A Fast Sequence contains the following timing settings
for sequence:
Setting
Setting Range
Description
Memo
12 characters
A user-created note for the
currently selected sequence.
Mode
CC, CR
Operating mode for the sequence.
Range
ILVL
Low I range, Low V range
IMVL
Middle I range, Low V range
IHVL
High I range, Low V range
ILVH
Low I range, High V range
IMVH
Middle I range, High V range
IHVH
High I range, High V range
Loop
Infinity,
01 ~ 9999
Sets the amount of times to loop
the selected sequence.
Last Load
OFF, ON
Set the load condition after the
end of the sequence.
Last
0.000000
A
The load setting for when Last
Load is set to ON.
RPTSTEP
0001 ~ 1000
Last step number (0001~1000)
per loop
Time Base
0.025 ~600ms
Sets the step execution time.