115
3-4-4. Results of OCP Test function
Power Source
OCP tripped
The OCP Test will return the current setting of the last
step when the power supply’s OCP was tripped.
OCP
RS232
01/Oct/2017
OCP Test
7.498
V
0.0000
A
0.00
W
Set Curr:
Fetch Volt:
2.000
1.500
1.000
6.91
6.92
6.91
Indicating power supply
OCP was tripped.
Current
:
Step current setting on the load
when the power supply
’
s OCP
was tripped.
2.500A
Power Source
OCP timeout
OC
P time out will occur if the power supply’s OCP fails to
trigger. This is determined when the measured voltage is
less than Trig V and the measured current is greater than
End C.
OCP
RS232
01/Oct/2017
OCP Test
7.498
V
0.0000
A
0.00
W
Set Curr:
Fetch Volt:
2.000
1.500
1.000
6.91
6.92
6.91
OCP time out
OCP time out indicates that the
power supply OCP was not
tripped.