2-1-8. Shorting the Load
Description
The Short key can be used to simulate a short circuit of
the load input terminals. A short circuit is simulated by:
Setting the current to the maximum value in CC mode.
Setting the resistance to the minimum value in CR
mode.
Setting the voltage to the minimum value in CV mode.
Setting the power to the maximum value in CP mode.
When the load is shorted, the external controller also
sends a short signal. See page 144 (
4-1-15. Short Control
)
for usage details.
Operation
The short function can be turned on and off by pressing
the
Short
key.
The
Short
key will turn red when the short function is
active.
The Short icon will appear when the short function is
active.
Display
0.100
V
0.10
W
LOAD
RS232
SHORT
01/Oct/2017
SHORT on
Note
If the load is already off, pressing the Short key will turn
the load on (shorted) at the same time.
Pressing the Short key again will also turn the load off
again as well.
If the load is already on and the Short key is pressed,
then when the Short key is pressed again the load will
remain on (the electronic load will return to its previous
load condition).
The Short key will be disabled if the Short Function
setting is turned off. See page 45 (
2-1-11.Short Function
Enable/Disable
) for details.
Summary of Contents for LSG-1050H
Page 24: ...14 GRA 414 E EIA standard 132 382 5 477 6 GRA 414 J JIS standard 149 382 5 477 6...
Page 185: ...175 7 6 Operating Area 7 6 1 LSG 175H H Range M Range L Range...
Page 186: ...176 7 6 2 LSG 350H H Range M Range L Range...
Page 187: ...177 7 6 3 LSG 1050H H Range M Range L Range...
Page 188: ...178 7 6 4 LSG 2100H H Range M Range L Range...
Page 204: ......