114
3-4-3. Running a OCP Test function
Operation
1.
Press
FUNC
>
OCP
[F4] >
OCP ON
[F1] in order.
2.
Turn the load on.
See page 87 (
3-1-2.Turning on the Load with the Selected
Function
) for the load on.
The test current will increase from the Start C value to
the End C value in steps according to the Step C
value, until the test has finished.
The test will start running when the power supply
voltage is greater than the Trig V voltage.
Example:
OCP Function
running
OCP
RS232
01/Oct/2017
OCP Test
7.498
V
0.1531
A
1.15
W
Set Curr:
Fetch Volt:
2.000
1.500
1.000
6.91
6.92
6.91
Measured voltage for
last three steps
Measured voltage,
current and power
Running.
Set current for the
last three steps
(descending order)