68
2-5-6. UnReg
Description
The UnReg error message will appear on the LCD
display when the electronic load is operating in an
unregulated state.
Alarm
The UnReg indicator will appear on the LCD display
when the set load is inadequate for the power source.
To clear the UnReg indicator, increase the current of
power source or reduce the load requirements.
Display
8.653
V
50.006
mA
0.4327
W
CV B Value
80.000
CV A Value
80.000
LOAD
RS232
A Value
Fine
01/Oct/2017
V
V
UnReg
UnReg indicator
Mode
CV
I Range
L87.5mA
V Range
L 80V
Response
Fast
Configure
2-5-7. Para
Description
The Para error message will appear on the LCD display
when the LSG-H is used in parallel and if an error is
produced.
Alarm
The Para error message indicates one of the following
possible conditions: UnReg, R.OCP, OTP.
To clear the Para indicator, remove the cause of the
alarm.
Display
8.653
V
50.006
mA
0.4327
W
CV B Value
80.000
CV A Value
80.000
LOAD
RS232
A Value
Fine
01/Oct/2017
V
V
Para
Para indicator
Mode
CV
I Range
L87.5mA
V Range
L 80V
Response
Fast
Configure
Summary of Contents for LSG-1050H
Page 24: ...14 GRA 414 E EIA standard 132 382 5 477 6 GRA 414 J JIS standard 149 382 5 477 6...
Page 185: ...175 7 6 Operating Area 7 6 1 LSG 175H H Range M Range L Range...
Page 186: ...176 7 6 2 LSG 350H H Range M Range L Range...
Page 187: ...177 7 6 3 LSG 1050H H Range M Range L Range...
Page 188: ...178 7 6 4 LSG 2100H H Range M Range L Range...
Page 204: ......