174
7-5-4. CV Mode
CV Mode
When the unit is set to CV mode it will operate as a constant
voltage operation when connected to a constant current
source.
This means the unit will maintain the CV setting value, up to
the rated power, regardless of the input current.
This is illustrated below.
Note that when the load voltage is less than the +CV setting
value, no current will flow due to very high impedance.
Load
current
Voltage
Current
Rated power level
Constant voltage operation
Rated current
Rated voltage
Maximum current
in input possible
CV setting
value
No current area