47
2-2.
Basic Configuration
The basic configuration settings are the common configuration settings that
are used for each operating mode. After selecting a basic operating mode (CC,
CR, CV and CP mode), the slew rate, switching function, response rate and
other common parameters should be configured.
2-2-1. Select the Switching Function
Description
The LSG-H Series has two switching function, static
mode and dynamic mode. The switching function allows
the LSG-H Series to switch between two preset levels.
Static mode can only switch between the two levels
manually, while Dynamic mode switches between each
level automatically based on a timer.
Static mode: A Value, B Value
Dynamic mode: Level1, Level2
When the unit is set to static mode, only one value (A
Value or B Value) can be active at a time. The active
value is shown in the operation status panel.
Function
Static
Configure
0.00
W
LOAD
RS232
A Value
Fine
W
W
Active
setting
When the unit is set to dynamic mode, the unit will switch
between Level1 and Level2 based on the Timer1 and
Timer2 parameters, shown below.
Note
Dynamic mode is not available for CV mode.
Time
Value
Dynamic mode
Timer1 Timer2
Level1
Level2
Timer1