153
5-1-2. Configure GP-IB Interface
To use GP-IB, the optional GP-IB port must be installed. See page 162 for
installation details (
7-2.GP-IB Installation
).
Operation
1.
Ensure the LSG-H Series is off before proceeding.
2.
Connect a GP-IB cable from a GP-IB controller to the
GP-IB port on the LSG-H Series.
3.
Turn the LSG-H Series on.
4.
Press
Shift
>
Utility
Help
>
Interface
[F3] in order,
and set the
Interface setting to
GP-IB
.
5.
Set the GP-IB address.
GP-IB address
0~30
GP-IB
constraints
Maximum 15 devices altogether, 20m cable length, 2m
between each device.
Unique address assigned to each device.
At least 2/3 of the devices turned On.
No loop or parallel connection.
Pin Assignment
1
12
13
24
Pin
Signal
Pin
Signal
1-4
Data I/O 1-4
13-16
Data I/O 5-8
5
EOI
17
REN
6
DAV
18
Ground (DAV)
7
NRFD
19
Ground (NRFD)
8
NDAC
20
Ground (NDAC)
9
IFC
21
Ground (IFC)
10
SRQ
22
Ground (SRQ)
11
ATN
23
Ground (ATN)
12
SHIELD Ground
24
Single GND