58
2-3-3. Timer Functions
2-3-3-1. Count Time
Description
When Count Time timer is set to on, it will count the
elapsed time from when the load was turned on to when it
was turned off.
This function is applicable to manual and automatic
shutdown (such as from protection functions such as
UVP etc.)
The elapsed time will be shown in the display
Measurement area.
Operation
Press
Main
>
Configure
[F5] >
Other
[F2] in order,
and turn the
Count Time
on or off.
Count Time:
ON, OFF
Display
0.000
V
0.0000
A
0.00
W
Level1
0.00
LOAD
RS232
01/Oct/2017
W
0:00:05
Elapsed time
2-3-3-2. Cut Off Time
Description
The Cut Off Time function will turn the load off after a set-
amount of time. After the load has been turned off, a
popup screen will LCD display the voltage level when the
load was turned off.
Operation
Press
Main
>
Configure
[F5] >
Other
[F2] in order,
and set the
Cut Off Time
.
Cut Off Time:
OFF, 1 s ~ 999 h 59m 59 s
Display
Mode
CP
I Range
H 8.75A
V Range
L 80V
Function
Dynamic
Configure
0.000
V
0.0000
A
0.00
W
Timer1
0.025
Level2
0.00
Level1
0.00
LOAD
RS232
Fine
01/Oct/2017
mS
W
W
0:00:05
Cut off time
Time Up
Voltage : 5.1223V
Enter
Voltage at cut off time
Summary of Contents for LSG-1050H
Page 24: ...14 GRA 414 E EIA standard 132 382 5 477 6 GRA 414 J JIS standard 149 382 5 477 6...
Page 185: ...175 7 6 Operating Area 7 6 1 LSG 175H H Range M Range L Range...
Page 186: ...176 7 6 2 LSG 350H H Range M Range L Range...
Page 187: ...177 7 6 3 LSG 1050H H Range M Range L Range...
Page 188: ...178 7 6 4 LSG 2100H H Range M Range L Range...
Page 204: ......