73
2-7.
Go-NoGo
The Go-NoGo configuration is used to create pass/fail limits on the voltage or
current input. If the voltage/current exceeds the pass/fail limits, an alarm will be
output.
The Go-NoGo configuration can be used with the program operation to create
complex pass/fail tests.
2-7-1. Setting the Go-NoGo Limits
Description
The Go-NoGo setting limits can be set as either discrete high
& low values or as a percentage offset from a center value.
The limit level of the CC, CR and CP mode become the
voltage level.
The limit level of the CV mode becomes the current level.
The set range of the limit level of the voltage/ current is rated
voltage/ current of the voltage/ current range H.
Operation
1.
Press
Main
>
Configure
[F5] >
Go-NoGo
[F3] in order.
2.
Select
Entry Mode
and choose how to set
the pass/fail limits.
Value will allow you to set the limits as discrete values.
Percent will allow you to set the limits as a percentage
offset from a center value.
3.
If
Entry Mode
was set to
Value
,
Set the
High
&
Low
limit values.
High:
0~ rated current/voltage
Low:
0 ~ rated current/voltage
4.
If
Entry Mode
was set to
Percent
,
Set the
Center
voltage/current and
High, Low
% values.
Center:
0~ rated current/voltage
High:
0~100% of Center current/voltage
Low:
Center - 0~100% of Center current/voltage
5.
Set the
Delay Time
.
The delay time setting will delay activating the Go-NoGo
testing by a specified amount of time.
The delay setting can compensate for startup oscillation
and other instabilities during startup.
Delay Time
0.0~1.0 seconds (0.1s resolution )
Note
When the Main settings are saved or recalled, the Go-NoGo
settings are also saved / recalled. See the Save/Recall
chapter for details, page 75 (
2-8. Save / Recall
).
Summary of Contents for LSG-1050H
Page 24: ...14 GRA 414 E EIA standard 132 382 5 477 6 GRA 414 J JIS standard 149 382 5 477 6...
Page 185: ...175 7 6 Operating Area 7 6 1 LSG 175H H Range M Range L Range...
Page 186: ...176 7 6 2 LSG 350H H Range M Range L Range...
Page 187: ...177 7 6 3 LSG 1050H H Range M Range L Range...
Page 188: ...178 7 6 4 LSG 2100H H Range M Range L Range...
Page 204: ......