6. FAQ
• The load voltage indicated on the load module is below expected.
• The front panel keys are not working.
• The load won’t turn on.
• The performance does not match the specification
The load voltage indicated on the load module is below expected.
Ensure the load leads are as short as possible, twisted and use the
appropriate wire gauge. Ensure that remote sense is used, this can help
alleviate the voltage drop across the load the leads.
The front panel keys are not working.
Check to make sure that the key lock has not been activated. LOCK will be
shown on the panel when the screen is locked. Press Shift + Clear (Lock) to
unlock the keys.
The load won’t turn on.
If you are using the load key to try to turn the load on and the load won’t turn
on, it is possible that external control is activated and that the LoadOn In
setting is set to low. See page 140 (
4-1-8.Turning the Load On using External
Control
) for details.
The performance does not match the specification.
Make sure the device is powered on for at least 30 minutes, within
+20°C~+30°C. This is necessary to stabilize the unit to match the specification.
For more information, contact your local dealer or
TEXIO TECHNOLOGY at