BATT Test function
Description
The BATT test function creates an automatic test to test the
discharge of Battery products.
The test will discharge in a fixed mode (CC, CR, CP) and will
end after a defined stop point (stop voltage, stop time, stop
AH) has been detected. The information about discharge test
(discharge time, battery AH, battery WH) can be finally seen
on the panel.
The PEL-3000(H) also has a user-defined cutoff setting in the
event that the Battery test fails.
The diagram below shows an example of the BATT Test function:
The test will run in the specified mode with defined values and will stop when
the defined stop values are reached.
Example:
Discharge CC mode
Vin
Current
t
SlewRate
=
0
Setting
CC mode
SlewRate
?
Stop Time
Begin Discharge
Stop AH
Stop Volt
AH