112
3-4-1. OCP Test function setting parameters
Parameters
No.
Selects one of 12 OCP test setup memories.
Memo
A user-created note for the currently
selected OPP function.
Range
Select the Range of CC Mode.
(High, Middle, Low)
Start Current
(Start C)
Starting start current value for the test.
End Current
(End C)
The current value that will end the test.
The value must be higher than the OCP
value of the power supply you are testing.
This parameter is used as a fail-safe for if the
over current protection of the power supply
fails.
If the measured current is reaches End
Current value it would then indicate that the
power supply OCP failed.
Step Current
(Step C)
Sets the step resolution of the current.
Step Time
(Step T)
Sets the execution time of each step. (50ms
to 1600s)
Delay Time
(Delay)
The OCP testing delay time.
Sets the how long to delay starting the test
after the Load On key has been pressed
(5ms ~ 160ms).
Trig Voltage
(Trig V)
Sets the trigger to a level needed to see
when the power supply OCP has been
triggered.
When the power supply OCP has been
triggered, its voltage output will reset.
The voltage trigger level is used to test to
see if the voltage output has been reset.
Last Current
(last C)
Sets the final current value after OCP has
been tripped.
This is the steady-state current draw after the
OCP has been tripped.
Note
This mode can only be used under CC mode.
Summary of Contents for LSG-1050H
Page 24: ...14 GRA 414 E EIA standard 132 382 5 477 6 GRA 414 J JIS standard 149 382 5 477 6...
Page 185: ...175 7 6 Operating Area 7 6 1 LSG 175H H Range M Range L Range...
Page 186: ...176 7 6 2 LSG 350H H Range M Range L Range...
Page 187: ...177 7 6 3 LSG 1050H H Range M Range L Range...
Page 188: ...178 7 6 4 LSG 2100H H Range M Range L Range...
Page 204: ......