74
2-7-2. Running a Go-NoGo Test
Description
Go-NoGo test results are displayed in the measurement
panel of LCD display.
GO indicates pass (good).
NG indicates fail (no good).
Operation
1.
Press
Main
>
Configure
[F5] >
Go-NoGo
[F3] in
order.
2.
Set
SPEC Test
to ON.
When SPEC Test is ON, SPEC will appear in the
operation status panel of LCD display. This means the
unit is ready for Go-NoGo testing.
3.
Turn the load on.
The test starts from the time the load was turned on +
the Delay Time.
Display:
GO
0.000
V
0.0000
A
0.00
W
Level2
0.00
Level1
0.00
LOAD
RS232
01/Oct/2017
W
W
SPEC test = ON
SPEC
GO
Test result
Display:
NG
0.000
V
0.0000
A
0.00
W
Level2
0.00
Level1
0.00
LOAD
RS232
01/Oct/2017
W
W
SPEC test = ON
SPEC
NG
Test result