3-5-3. Running a OPP Test function
Operation
1.
Press
FUNC
>
Next Manu
[F5]
> OPP
[F1] in
order,
and Press
OPP ON
[F1] to turn OPP ON.
2.
Turn the load on.
See page 87 (
3-1-2.Turning on the Load with the Selected
Function
) for the load on.
The test watt will increase from the Start W value to
the End W value in steps according to the Step W
value, until the test has finished.
The test will start running when the power supply
voltage is greater than the Trig V voltage.
Example:
OPP Function
running
Exit
OPP
USB
01/Oct/2018
OPP Test
4.88
V
4.142
A
W
50.61
Fetch Watt: Fetch Volt:
10.03 4.91
0.00 4.92
Running.
10.03