95
3-2-4. Running the Program function
Description
Turn the road on, the program function is running.
Operation
1.
Press
FUNC
>
Program
[F1] in order.
2.
Turn program mode on by setting
Program
[F1] to on.
PROG
will appear at the top of the LCD display when
Program
is On.
3.
Turn the load on.
See page 87 (
3-1-2.Turning on the Load with the Selected
Function
) for the load on.
The
PROG
icon turns orange when the load is turned
on.
4.
When a program is running the screen displays which
program, step and memory is currently active.
Press
Pause
[F1] to suspend a test, press
Continue
[F1] to resume.
Press
Next
[F2] to run the next step if its
Run
setting
was set to
Manual
.
5.
When a program has finished running, a list of the Go-
NoGo results for each step are displayed.
Press
Exit
[F5] to exit.
Display:
Program
Running
Continue
PROG
RS232
01/Oct/2017
Run Program
0.000
V
0.0000
A
0.00
W
Program No:
Step(Memory)
01(001)
01
GO
Program number that
is currently running.
Go-NoGo result
for the step
Memory number
of current step.
Step that is
currently running.
Display:
Program
Finished
Exit
PROG
RS232
01/Oct/2017
Run Program Detail Result
1
1
1
Program
3
2
1
Step
NG
GO
GO
Result
Summary of Contents for LSG-1050H
Page 24: ...14 GRA 414 E EIA standard 132 382 5 477 6 GRA 414 J JIS standard 149 382 5 477 6...
Page 185: ...175 7 6 Operating Area 7 6 1 LSG 175H H Range M Range L Range...
Page 186: ...176 7 6 2 LSG 350H H Range M Range L Range...
Page 187: ...177 7 6 3 LSG 1050H H Range M Range L Range...
Page 188: ...178 7 6 4 LSG 2100H H Range M Range L Range...
Page 204: ......