8
Parts Description
Screen Description
In the TCI mode
Graph screen
7
6
5
4
1
2
3
1
Drug display
Displays the drug name with a colour tag, the drug concentration, and
the model.
2
Tab display
Indicates the currently selected target (plasma or effect-site targeting)
and the type of TCI mode screen (graph, text, or info).
3
Target concentration
display
Displays the predicted concentration (Cpt or Cet) to be targeted.
4
Solution delivery
information display
Displays the current flow rate, dose rate, decrement (awakening) time,
and volume delivered.
5
Graph display
Displays the change in a time lapse of the predicted concentration (Cp
or Ce) in a graphical format. The lines/areas are displayed in different
colours that are determined by the selected drug.
6
Cp/Ce display
Displays the current value of Cp (predicted plasma concentration) and
Ce (predicted effect-site concentration).
7
Initial infusion display
Displays the initial infusion dose and the time required before the target
concentration (Cpt or Cet) level is reached. (Available before the initial
infusion is started.)
Notes
• Each time you press the Display select switch, the screen changes as shown below.
• The display returns to the graph screen after 10 seconds of no operation on the Info screen.
Graph screen
Text screen
Info screen