36
Operation Procedure
Confirmation switch
Increases value
Dial
Decreases
value
4)
Turn the Dial to set the gender, age, height, and weight, then
press the Confirmation switch.
Notes
• The setting range is as follows:
Age:
12 to 100 (Remifentanil), 16 to 100 (Propofol)
Height: 100 to 250 cm
Weight: 30.0 to 200.0 kg
• The default setting is Male, 40 Years Old, 170 cm, and 70.0 kg.
• When Propofol (Marsh) is selected, “Gender” and “Height” are not displayed.
Confirmation
switch
5)
Check the parameter value and that “OK?” is selected, then
press the Confirmation switch.
Note
• “Cpt” is displayed instead of “Cet” when plasma TCI is selected.
6)
Check
The selected drug name, model, and target are displayed,
and the syringe displacement icon
is not displayed.
Warning
• Before using the TCI mode, make sure the correct drug name, model, target,
and target concentration are displayed. Incorrect settings of drug name,
model, target, and target concentration may cause significant health hazard to
the patient.
Note
• The colour tag and the graph lines/areas are displayed in unique colours for each
drug, which makes it easier to identify the currently selected drug.
Target
Drug name with colour tag
Model