45
Operation Procedure
Increases value
Dial
Decreases
value
6)
Turn the Dial to change the scale of the Y axis (plasma or
effect-site concentration).
Notes
• The scale of the Y axis can be selected from AUTO, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20,
50, 100, 200, 500, 1000, and 2000.
• The default setting of the Y axis is AUTO.
• When AUTO is selected, the suitable scale is automatically selected so that the
entire graph is displayed on the screen. A bar appears above the number indicating
the currently selected scale.
Confirmation
switch
7)
Press the Confirmation switch.
Replacing the Syringe
Solution delivery can be continued by replacing the syringe with a new one.
Caution
• To continue the TCI infusion to the same patient, do not turn off the power while replacing the syringe with a new one.
Otherwise, all important information on the current TCI mode, such as the predicted concentration, will be erased.
1)
To continue solution delivery when the remaining amount in
the syringe is low, prepare a new one.
Stop switch
Operation indicator
2)
Press the Stop switch.
Notes
• The buzzer sounds and the operation indicator turns off, then solution delivery
stops.
• The predicted concentration remains displayed while the pump is stopped.
3)
Close the infusion line.
Note
• For details on stopping solution delivery, see page 61.
Summary of Contents for TERUFUSION TE-SS830P
Page 163: ...163 Memo ...