150
Specifications
Available
Functions
Power OFF buzzer
sound function
Buzzer sounds as the power is turned off.
Dose mode select
function
The dose mode can be selected among mL/h,
μ
g/kg/min, mg/kg/h,
mg/kg/h+DIPRIVAN, TCI, and drug library.
Start-up dose mode
select function
The initial dose mode after the power is turned on can be selected from mL/h,
μ
g/kg/min, mg/kg/h, mg/kg/h+DIPRIVAN, TCI, and drug library.
Dose mode information
display function
When the AC power cable is connected, the versions of TCI and drug library,
and profile name are displayed before the power is turned on.
Menu1 (User mode1)
shortcut function to
dose mode
Enable to go directly to the dose mode select display when the menu screen is
selected.
Drug/patient
information display
function
Displays drug information such as dilution, patient information including
patient name, age, gender, weight, height, etc. The dilution and weight can be
changed.
Dilution setting
function by amount and
solution
Selects how to specify the dilution from either Dilution or Amount/Solution.
Interval/Intermittent
dose function
The specified dosage rate is delivered at the specified regular intervals.
Multi-step dose
function
The speed of solution delivery increases or decreases at the specified levels
(up to 10 levels).
Delayed Start dose
function
Solution delivery starts after the specified time has elapsed.
Link function (Switch
function A, B)
Switch function A: Operations of two pumps are linked. When a Completion
alarm occurs on the first pump, the second pump starts
solution delivery.
Switch function B: Interlocks two pumps. Starts solution delivery of the first
pump, and then starts solution delivery of the second pump
at the time set in advance or afterwards. After solution
delivery of the second pump is started, each pump changes
the solution delivery speed according to respective settings.
The interlock of the pumps finishes upon completion of the
set solution delivery or at alarm timing.
Come and See ME
function
Automatic call function after a specified time elapses.
Volume delivered unit
select function
Switches the unit of volume delivered appearing in the drug library mode, the
μ
g/kg/min mode, the mg/kg/h mode, the mg/kg/h+DIPRIVAN mode, and the
TCI mode.
Micro notation select
function
Switches the unit between “
μ
g” and “mcg” appearing in the drug library mode,
the
μ
g/kg/min mode, the mg/kg/h mode, the mg/kg/h+DIPRIVAN mode, and the
TCI mode.
Graph display function
(TCI mode only)
Displays the chronological time-lapse change of TCI infusion in graphs.
The graph scale is adjusted accordingly as the time lapse when the relevant
mode is selected.
Calculated plasma
concentration display
function (TCI mode
only)
Displays the predicted plasma concentration of drug in the TCI mode.
Effect-site
concentration display
function (TCI mode
only)
Displays the predicted effect-site concentration of drug in the TCI mode.
Maximum flow rate
setting (TCI mode only)
The maximum flow rate in the TCI mode can be set.