160
For Medical Staff
Table b) – Manufacturer’s declaration – electromagnetic immunity – for IEC 60601-1-2:2014 (EN 60601-1-2:2015)
Manufacturer’s declaration - electromagnetic immunity
The TERUFUSION Syringe Pump Type SS3 OTCI is intended for use in the electromagnetic environment specified below. The
customer or the user of the TERUFUSION Syringe Pump Type SS3 OTCI should assure that it is used in such an environment.
Immunity test
Immunity test level
Electrostatic discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
Radiated RF
IEC 61000-4-3
3 V/m 80 MHz to 2.7 GHz
See Table c) for immunity test level to RF Wireless
communication equipment
Electrical fast transient / burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input / output lines
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Conducted RF
IEC 61000-4-6
3 Vrms 150 kHz to 80 MHz outside ISM bands
6 Vrms 150 kHz to 80 MHz in ISM bands
Power frequency (50-60 Hz) magnetic field
IEC 61000-4-8
30 A/m
Voltage dips, short interruptions and voltage variations on
power supply input lines
IEC 61000-4-11
0% U
T
for 0.5 cycle
0% U
T
for 1 cycle
70% U
T
for 25/30 cycles
0% U
T
for 250/300 cycles
Table c) – Immunity test level to proximity fields from RF communications equipment – for IEC 60601-1-2:2014 (EN 60601-1-2:2015)
Immunity test level to proximity fields from RF communications equipment
The TERUFUSION Syringe Pump Type SS3 OTCI is intended for use in the electromagnetic environment specified below. The
customer or the user of the TERUFUSION Syringe Pump Type SS3 OTCI should assure that it is used in such an environment.
Emissions test
Test frequency (MHz)
Immunity test level (V/m)
Proximity fields from RF communications
equipment
IEC 61000-4-3
385
27
450
28
710, 745, 780
9
810, 870, 930
28
1720, 1845, 1970
28
2100
10
2450
28
2600, 3500
10
5240, 5500, 5785
9