135
Troubleshooting
High priority alarm
Plunger Displacement alarm
Occurrence
Cause
Action
• LCD: The Plunger
Displacement alarm display
flashes red.
• Operation indicator: Flashes
red.
• Buzzer: Activates.
• Pump: Stops.
• The plunger has been
displaced during solution
delivery.
1. Re-set the syringe correctly.
(See pages 32 and 51.)
2. Prime the line, ensuring that the plunger is kept in the
slider to prevent the solution from being delivered to
the patient. (See pages 38 and 58.)
• LCD: The Plunger
Displacement alarm display
flashes red.
• Operation indicator: Flashes
red.
• Buzzer: Activates.
• Pump: Remains stopped.
• Unable to start even after the
syringe has been correctly set.
• Drug solution is adhered on
the plunger detector.
• The malfunction of the internal
sensor of the plunger detector
is suspected.
(After the alarm is issued, the
solution delivery cannot start
while the power is on.)
• If any adhered drug solution is present, turn off the
power and clean the plunger detector.
• If the problem persists, contact TERUMO trained
service technicians to request repair.
Syringe Displacement alarm
Occurrence
Cause
Action
• LCD: The Syringe
Displacement alarm display
flashes red.
• Operation indicator: Flashes
red.
• Buzzer: Activates.
• Pump: Stops.
• The syringe has been displaced during
solution delivery.
1. Replace the syringe correctly.
(See pages 32 and 51.)
2. Prime the line, ensuring that the
plunger is kept in the slider to prevent
the solution from being delivered to the
patient. (See pages 38 and 58.)
• LCD: The Syringe
Displacement alarm display
flashes red.
• Operation indicator: Flashes
red.
• Buzzer: Activates.
• Pump: Remains stopped.
• Unable to start even after the
syringe has been correctly set.
• Drug solution is adhered on the syringe
clamp and/or flange holder, or they are
damaged.
• The malfunction of the internal sensor of
the syringe clamp and/or flange holder is
suspected.
(After the alarm is issued, the solution
delivery cannot start while the power is on.)
• If any adhered drug solution is present,
turn off the power and clean the syringe
clamp and/or flange holder.
(See page 129.)
• If the problem persists, contact TERUMO
trained service technicians to request
repair.