13
Parts Description
Occlusion icon
The internal pressure level of the
infusion line that does not reach
the level of Occlusion alarm is
displayed in 4 levels.
(grey)
Below 25% of the specified occlusion detection pressure.
(green)
25% or more of the specified occlusion detection pressure.
(yellow)
50% or more of the specified occlusion detection pressure.
(orange)
75% or more of the specified occlusion detection pressure.
Syringe brand/Syringe size
icon
Syringe brand
Syringe size to be used (5, 10, 20, 30, 50)
(If a syringe has not been set or has not been correctly detected, the size
display indicates “--”. )
Keypad lock icon
Displays when the keypad lock is
enabled (ON).
When the keypad lock is enabled (ON).
(No indicator)
When the keypad lock is disabled (OFF).
Mute icon
Displays when an alarm is muted.
Alarm is muted
Maintenance icon
Displays when the maintenance
timer set period has elapsed.
When the set time period has elapsed
For the setting method, contact TERUMO trained service technicians.
Alarm icon
Displays when an alarm sounds.
(Background
colour: red)
When a high priority alarm sounds.
(Background
colour: yellow)
When a medium priority alarm sounds.
(Background
colour: yellow)
When a low priority alarm sounds.
Wireless LAN icon
Displays the strength of the
signal.
Antenna only is flashing:
The wireless LAN module
is faulty.
Antenna only is lit:
Connection to the network
system is not established.
A signal strength bars light grey:
Connection to the network system is
established.
A signal strength bars light green:
During communication.
Signal condition: Poor
Signal condition: Strong
Communication icon
Displays during the infrared
communication (IrDA
®
).
During communication