Operation Procedure
Setting Flow Rate
Set the flow rate per hour.
Increases value
Dial
Decreases
value
•
Use the Dial to adjust the flow rate in the flow rate screen.
Note
• The flow rate screen appears in dark blue (selected).
Start up icon
Notes
• The flow rate can be set within the range of the following values.
Setting
range:
0.01 to 150.00 mL/h
Note that the upper limit of the flow rate can be changed in the following ranges:
0.01 to 150.00 mL/h (when using syringe of 5 mL)
0.01 to 300.00 mL/h (when using syringe of 10 mL, 20 mL, 30 mL)
0.01 to 1200.00 mL/h (when using syringe of 50/60 mL)
Step:
0.01 mL/h step (0.01 to 10.00 mL/h)
0.10 mL/h step (10.00 to 100.00 mL/h)
1.00 mL/h step (100.00 to 1200.00 mL/h)
• When 150.00 mL/h is reached, the confirmation message “This is the hard upper limit” is displayed on the LCD.
• For safety, the Dial is designed to make the initial half turn without changing the value on the display. When the value starts
to change, the buzzer sounds momentarily for confirmation. When you stop rotating the Dial and start it again in less than 1.5
seconds, the value changes without the initial half turn.
• Rotating the Dial while pressing the Stop switch allows changing the setting at the rate of 100 times.
• When the Dial is turned, a line appears under the numerals to indicate the currently changing digit. (When the Dial is rotated
while pressing the Stop switch, an underline appears to the second digit to the left from the normal digit.)
• The set flow rate can be changed during solution delivery. When the Dial is turned on the flow rate screen during solution
delivery, the flow rate setting area turns dark blue where the set flow rate increases or decreases in accordance with the Dial
operation. The Start switch confirms the set flow rate and changes the solution flow rate. (When the Start switch is not pressed
within approx. 5 seconds, it returns to the set flow rate before change.)