31
Operation Procedure
Using the TCI mode
Turning the Power On
Please ensure the product is fit for purpose prior to use.
The TCI mode is an alternative that enhances the convenience and control of intravenous anesthetic delivery.
The pump calculates and controls the flow rate to achieve and maintain the target concentration (Cpt or Cet). For details, see page
3.
Cautions
• Conduct the self-check (self-diagnosis) of this product without a syringe installed. Otherwise the self-check may not be
conducted properly.
• Check for any fault by conducting a self-check. If a fault is observed, immediately discontinue use and contact TERUMO
trained service technicians.
1 second
or more
•
Press and hold the Power switch for 1 second or more until
the TERUMO logo screen is displayed to turn the power on.
Checks
When turning the power on, the following occur simultaneously;
• TERUMO logo appears on the display.
• LCD flashes 3 times.
• Operation indicator repeats lighting green and red alternately, and then buzzer sounds.
Operation indicator
Notes
• A timer for maintenance can be set in this product. When the period set for the
maintenance timer is reached, a maintenance request is displayed at start up.
When the request is displayed, contact TERUMO trained service technicians.
• When the period set for the maintenance timer is reached, this screen displays
at every start up until the setting is changed. For the setting method, contact
TERUMO trained service technicians.