76
Other Operation Procedures
Confirmation
switch
4)
Press the Confirmation switch.
Notes
• When
μ
g/kg/min or mg/kg/h is selected: The input screen of amount and solution
appears.
• When mg/kg/h+DIPRIVAN is selected: The input screen of body weight appears
since dilution is fixed to 10 mg/mL. Go to step 6.
Confirmation switch
Increases value
Dial
Decreases
value
5)
Use the Dial to adjust the value of amount and solution, and
press the Confirmation switch.
Notes
• The amount and solution can be set within the range below.
Amount (drug amount):
0.01 to 999.00 mg*
Solution (solution amount): 0.01 to 999.00 mL*
* Setting range of the dilution = Amount : 0.01 to 999.00 mg/mL
Solution
• If you want to set by Dilution, contact TERUMO trained service technicians.
• If the Confirmation switch is pressed after setting the drug amount, the input area
moves to Solution.
• When the Confirmation switch is pressed after setting the solution amount, the
input screen of body weight appears.
• When the Back/Mute switch is pressed, the parameter input is cancelled and the
screen returns to the previous status.
Confirmation switch
Increases value
Dial
Decreases
value
6)
Use the Dial to adjust the value of weight, and press the
Confirmation switch.
Notes
• Weight can be set within the range of 0.1 to 300.0 kg.
• Pressing the Confirmation switch displays the parameter confirmation screen.