138
Troubleshooting
Shutdown Notice alarm
Occurrence
Cause
Action
• LCD: The Shutdown Notice
alarm display flashes red.
• Operation indicator: Flashes
red.
• Buzzer: Activates.
• Pump: Stops.
• Battery has no power left.
• Connect immediately to the AC power supply or rack
system. (See page 29.)
* When using a rack system, see the instruction manual
of the rack system.
Power Failure alarm
Occurrence
Cause
Action
• LCD: The Power Failure alarm
display flashes red.
• Operation indicator: Flashes
red.
• Buzzer: Activates.
• Pump: Stops.
Two of the following causes have
occurred.
• No AC power supply
• Battery failure
• Sub-battery failure or low
battery
• Connect to the AC power supply and charge (8 hours or
more) with the power turned off.
If the problem still persists, contact TERUMO trained
service technicians.