Changing Settings on the Menu
Confirmation
switch
5)
Press the Confirmation switch.
Notes
• The bolus setting display appears.
• When the flow rate screen is displayed, the Hands Free Bolus screen can be
displayed by pressing and holding the Back/Mute switch and pressing the Purge
switch twice.
Increases value
Dial
Decreases
value
6)
Turn the Dial to set the Bolus VTBI.
Notes
• The Bolus VTBI can be set within the range below.
VTBI: 0.01 to 999.00
• The Bolus VTBI must be set within the values obtained by subtracting the volume
delivered from the VTBI.
Display
select switch
7)
Press the Display select switch to select the Bolus VTBI
time or the Bolus flow rate.
Bolus flow rate
Bolus VTBI time