139
Troubleshooting
Medium priority alarm
Pressure alarm
Occurrence
Cause
Action
• LCD: The Pressure alarm
display flashes yellow.
• Operation indicator: Flashes
green.
• Buzzer: Activates.
• Pump: Delivers solution.
• The internal pressure of
the infusion line has rapidly
increased or decreased.
• Due to a possible fault in the infusion line or at the
puncture site, check for any fault. Stop the solution
delivery and eliminate the cause.
Note
• This alarm is effective only when the Pressure alarm has been set in the drug library.
Low priority alarm
No Flow Rate alarm
Occurrence
Cause
Action
• LCD: The No Flow Rate alarm
display lights up yellow.
• Operation indicator: Off.
• Buzzer: Activates.
• Pump: Remains stopped.
• The Start switch was pressed
without having set a flow rate.
• Set a flow rate and press the Start switch to start
solution delivery. (See page 55.)
• The flow rate setting exceeds
150 mL/h (for 5 mL syringe) or
300 mL/h (for 10, 20 or 30 mL
syringe).
• Check the value set and re-set it to equal or less than
150 mL/h or 300 mL/h, depending on the syringe size
used. (See page 55.)
• Some of the flow rate settings
do not apply to the advanced
dose or combination dose
mode.
• Correct those settings and start solution delivery.
(See page 96. For the combination dose mode
setting, contact TERUMO trained service technicians.)
• In the TCI mode, the target
concentration (Cpt or Cet) was
set to 0.0 and the Start switch
was pressed. (Except during
solution delivery)
• Set the target concentration (Cpt or Cet) in the TCI
mode, then press the Start switch. (See page 35.)