131
After Use
Storage
After inspecting the pump, store it in an environment which conforms to the following storage conditions.
Temperature:
-20°C to 45°C
Humidity:
10% to 95%RH (non-condensing)
Atmospheric pressure: 50 to 106kPa (500 to 1060hPa)
Ensure that the battery is charged in preparation for it’s next use.
Warnings
• Since this product does not have an airtight structure, it should not be used or stored in an active gas environment
(including sterilizer gas), nebulizer-sprayed environment, high-humidity environment, etc. It should not be submerged
into water. If the electronic components inside the product are affected, there may be subsequent damage and time
degradation which will cause failure of this product.
• The product should not be used or stored in a flammable environment.
Cautions
• Do not store this product in a place with a high level of vibration, dust, mist or corrosive gas.
• Do not expose this product to direct sunlight or ultraviolet irradiation for a long period of time. The exterior may experience
colour change, deformation or deterioration.
• Do not store this product in a place where an environment of atmospheric pressure, temperature, humidity, ventilation, salt
content or sulfur content may cause an adverse effect.
• Do not store this product in a storage area for chemicals or in a place that generates gases.
• When transporting this product, avoid any shock, vibration, dust, high temperature and high humidity. Transport conditions:
Temperature: -20 to 60°C, Humidity: 10 to 95%RH (non-condensing), Atmospheric pressure: 50 to 106kPa (500 to
1060hPa)
• If the battery is not used for a long period after being charged, the amount of charge decreases due to self-discharge. Do
not store the battery in a discharge status. Storing the battery left in a discharge status may cause degradation, resulting in
making it unavailable in emergency situations.
Summary of Contents for TERUFUSION TE-SS830P
Page 163: ...163 Memo ...