73
Other Operation Procedures
Setting Special Functions
The following settings are selectable by changing the internal settings of this product. Taking account of a subsequent impact on
product performance, contact TERUMO trained service technicians.
Setting item
Description of function
VTBI setting
Allows you to set the VTBI. (For details, see page 56.)
VTBI time setting
Allows you to set the time limit. (For details, see page 57.)
Setting dial acceleration function
Rotating the Dial while pressing the Stop switch accelerates the setting rate to set items such
as the flow rate. This function is not available during solution delivery. Pressing the Stop
switch stops the solution delivery.
• Flow rate and VTBI: change the rate to 100-step increments
• Date and time: change the rate to hour and month
• History and library: change the rate to page feeding
Syringe brand setting
The syringe brand can be selected after the power is turned on and the syringe installation
has finished.
Flow rate range setting function A
(Soft limits)
Flow rate range setting function B
(Hard limits)
Sets the limitation of the flow rate in 2 levels; the soft limit (can be exceeded after
confirmation) and the hard limit (cannot be exceeded).
• When setting the flow rate over the soft limit, the flow rate will be displayed in orange.
• When the drug library mode or body weight dose mode (
μ
g/kg/min, mg/kg/h, or
mg/kg/h+DIPRIVAN) is selected, the flow rate range setting function is not available.
(See page 66.)
Change the Nearly Empty alarm
setting
Allows you to set the Nearly Empty alarm position based on the time before the syringe is
empty.
Change the KVO flow rate
Allows you to set the KVO flow rate within the range of 0.10 to 3.00 mL/h.
Hands Free Bolus setting
When bolus volume and bolus time are set, flow rate is calculated and the bolus infusion is
available. (For details, see page 88.)
Volume delivered unit select
function
In the case of mass system dose mode, the unit for volume delivered or VTBI can be
changed from volume (mL) to mass (g).
Alarm melody select function
Changes the alarm melody from Pattern 1 to Pattern 2.
The default setting is Pattern 1.
Night mode setting
If the device is left non-operated for 30 seconds between 21:00 to 5:00 (customizable), the
brightness can be automatically lowered by two levels (customizable).
Change the standby duration
Up to 24 hours with 10 minute intervals.
Dose mode select function
The dose mode can be selected among mL/h,
μ
g/kg/min, mg/kg/h, mg/kg/h+DIPRIVAN, TCI,
and drug library.
Start-up dose mode select
function
The initial dose mode after the power is turned on can be selected.
Select from mL/h,
μ
g/kg/min, mg/kg/h, mg/kg/h+DIPRIVAN, TCI, and drug library.
Dilution setting function by
amount and solution
Selects how to specify the dilution from either Dilution or Amount/Solution.
Micro notation select function
Switches the dose unit from “
μ
” to “mc”.
ON: mc, OFF:
μ
Selecting dose mode during the
solution delivery function
Change the dose mode from mL/h to another dose mode (except TCI) during solution
delivery.
The flow rate setting and volume delivered in the mL/h mode are retained after the dose
mode is changed.
• If the flow rate (mL/h) is smaller than the lower limit or bigger than the upper limit of the
hard limit in the setting value of the drug library etc., this function is not available.
Purge volume upper limit setting
function
The upper limit of purge volume can be set. The purging stops when the set upper limit is
reached.
Bolus volume upper limit setting
function
When the Hands On Bolus function is enabled, the upper limit of bolus volume can be set.
Bolus stops when the set upper limit is reached.
Menu1 (User mode1) shortcut
function to dose mode
Enable to display the dose mode select display by pressing and holding the Display select
switch for 2 seconds or more.
ON: Displays the dose mode select screen.
OFF: Displays the menu 1 screen.