128
In order to ensure the safe use of the syringe pump, make sure it is cleaned and inspected after use and stored in an appropriate
environment. If you find any damage or malfunction, immediately stop using the pump and contact TERUMO trained service
technicians.
Cleaning
Cautions
• Clean the AC inlet, slider and clutch on a regular basis. Failure to conduct the regular cleaning may result in damage,
failure or malfunction to/of this product.
• If any parts (clamp, clutch part etc.) of this product need to be removed for cleaning, contact TERUMO trained service
technicians.
• When disinfecting, do not use a sterilizer. Use a gauze (damp with antiseptic solution) to wipe the product, and wipe off the
antiseptic solution with a gauze (damp with cold/lukewarm water), and then thoroughly wipe off any moisture with a dry soft
cloth.
• Before cleaning, always turn off the power and disconnect the AC power cable. Otherwise it may cause an electrical shock
hazard. When AC power is supplied by the optional rack system (Model: TE-RS700, TE-RS800, TE-RS811), first remove
the pump from the rack system, then conduct cleaning.
• If any adhered drug solution is present, the solution delivery or alarm detection may not operate correctly or may interfere
with the normal operation of the clamp part and/or movable part of the pole clamp. If any drug solution is present, clean
immediately with a cotton swab.
• Do not wipe with undesignated alcohol, other organic solvents such as thinner, or povidone-iodine. Using organic solvent
or any antiseptic solution other than those permitted for use may result in damage to or failure of this product.
• Do not use an electric dryer to dry this product. This product may be damaged.
• This product and the pole clamp does not have a waterproof structure. Therefore, do not wash with running water or
immerse in water. It may cause damage or failure of the product.
• If the flange holder or clutch suffers from stiff movement or gets very dirty, it should be cleaned with a damp (not wet) soft
cloth.
Examples of disinfectants (ingredient names) which can be used are listed below.
Examples of Cleaning Disinfectants (Ingredient Names)
Ingredient Name
Dilution (e.g.)
Chlorhexidine gluconate
5%
Benzalkonium chloride
10%
Ethanol*
76.9 - 81.4vol%
When using disinfectant, follow the instruction manual of each disinfectant (regarding degree of dilution, etc.).
* Do not use ethanol to disinfect the following:
• One Touch Pole Clamp (specified parts only)
For details, see page 129.
After Use
Summary of Contents for TERUFUSION TE-SS830P
Page 163: ...163 Memo ...