143
Troubleshooting
Other problems
When installing a syringe: Plunger displaced
Occurrence
Cause
Action
• LCD: [Place the syringe.]
Slider displacement icon
flashes.
• The plunger has been
displaced.
• Place the syringe correctly. (See pages 32 and 51.)
When installing a syringe: Syringe displaced
Occurrence
Cause
Action
• LCD: [Place the syringe.]
Syringe displacement icon
flashes.
• The clamp is still open.
• The flange is not in the slit.
• The syringe is not installed
correctly.
• Place the syringe correctly. (See pages 32 and 51.)
• LCD: The Syringe
displacement icon
flashes.
The syringe brand setting is
incorrect.
• Check that the syringe brand setting matches the brand
of the syringe to be set.
When installing a syringe: Slider displacement
Occurrence
Cause
Action
• LCD: [Place the syringe.]
Slider displacement icon
flashes.
• The slider has been displaced. • Place the syringe correctly.
(See pages 32 and 51.)