47
Operation Procedure
Ending the TCI mode
Stop switch
Operation indicator
1)
Press the Stop switch.
Note
• The buzzer sounds and the operation indicator turns off, then solution delivery
stops.
Display
select switch
2 seconds
or more
2)
Press and hold the Display select switch for 2 seconds or
more.
Note
• The menu 1 screen (End of TCI screen) is displayed.
Confirmation
switch
3)
Press the Confirmation switch.
Note
• The display returns to the graph screen after 10 seconds of no operation.
Confirmation
switch
4)
Press the Confirmation switch to select End of TCI.
Note
• “Confirm end of TCI” is displayed on the LCD.