132
Maintenance and Inspection Items by TERUMO Certified Service Technicians
Cautions
• If any fault is observed, immediately stop using the product and contact TERUMO trained service technicians.
• Do not disassemble, make alterations (including actions that interfere with the functionality or performance such as taping
the LCD or a movable part) to, or repair this product. This may result in failure, damage or device performance degradation
of this product.
• Do not use any replacement parts other than those specified. The original functionality or performance of this product may
not be achieved.
Periodic Maintenance
Perform a periodic maintenance inspection to ensure safe operation and the longest possible life of the syringe pump. Contact your
local distributor for details on the nature and frequency of the maintenance inspection.
From TERUMO
• When ordering repairs or maintenance inspections, if there is a possibility of infection, disinfect the product in advance.
Battery replacement
Periodically replaced parts
Over a period of time, the battery will gradually deteriorate. Please see chart below for replacement schedule.
Part name
Elapsed years
Replacement reason
Battery (Lithium-ion battery)
2.5 to 3 years
• Even when charged, after a short period of time “Battery alarm”
is displayed and the buzzer sounds.
Sub-battery (Nickel-metal-hydride battery)
1 to 2 years
• The Power Failure alarm does not function properly.
* Depending on frequency of use and usage environment, the time for replacement for each part may change and replacement of
parts other than those scheduled for replacement may also be required. Consult TERUMO trained service technicians regarding
necessity and replacement.
* The life span of the equipment is at least 6 years in the case of standard use.
Fuse replacement
This product incorporates a fuse inside the structure. For replacement, contact TERUMO trained service technicians.
Waste and Recycle
Electrical and electronic equipment (EEE) and batteries contain materials, components and substances which can be dangerous to
the environment and harmful to human health if waste electrical and electronic equipment (WEEE) and batteries are not disposed
of correctly.
Waste electrical and electronic equipment and batteries must not be disposed of with the remainder of unseparated waste, but
should instead be collected separately. In this way, the environmental impact associated with disposal of WEEE and batteries is
reduced and there will be more opportunity for reusing, recycling and recovering WEEE and recycling batteries.
The Lithium-ion battery and Nickel-metal-hydride battery should be removed from the equipment by TERUMO certified service
technicians. Please contact your local distributor.
At end of life, please dispose of this equipment and batteries according to your local regulations. Contact your local distributor or
municipality for details of the available collection schemes. The embedded memory back-up lithium coin battery, collected together
with this equipment at end of life as WEEE, is to be removed and treated by the recycling center.
Maintenance and Inspections