33
Operation Procedure
Slider hook
Slider
Clutch
Plunger
Slider hook
4)
While holding the clutch, move the slider until it meets the
plunger, and then release the clutch.
If the following guidance in the figure below does not
display, perform this step again.
Checks
• The clutch appearing in the LCD should look as shown below.
• The slider displacement icon is turned off.
Note
• If the process is not completed after several attempts, the clutch may be ill fitting.
Ensure that the infusion line is not attached to a patient and then apply light
pressure to the slider to push the clutch into position.
The clutch may click when fitted into the position.
Clamp
2
1
5)
Turn the clamp and gently push it to fix the syringe.
Check
• The message “Syringe Barrel OK xxx mL” is displayed on the LCD, and the syringe
displacement icon
is not displayed. (xx is the size of the held syringe.)
Display select switch
Dial
6)
Turn the Dial or press the Display select switch to select the
correct syringe brand.
Warning
• After the syringe type is selected, recheck that the setting has been adjusted
correctly.
Notes
• Turning the Dial or pressing the Display select switch moves the white frame.
• The pump will automatically detect the syringe size.
Syringe brand tab
Summary of Contents for TERUFUSION TE-SS830P
Page 163: ...163 Memo ...