EMC Technical Information
Medical electrical equipment requires particular care in regards to EMC, and it is necessary to install and use the equipment in
accordance with the following EMC information.
Cautions
• Ensure that the power cable used is the one provided for this product. If not, this product’s emissions may increase and its
immunity may decrease.
• When using this product in combination with other equipment such as a network system, do not place the equipment next
to each other or stack them together. It may cause malfunctioning due to electromagnetic interference.
• Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the TERUFUSION Syringe Pump Type SS3 OTCI, including cables
specified in this manual. Otherwise, degradation of the performance of this equipment could result.
For Medical Staff