144
Troubleshooting
Maintenance timer
Occurrence
Cause
Action
• The Maintenance timer
indicator
lights up.
• It is time to perform
maintenance.
• Contact TERUMO trained service technicians.
AC power detector function
Occurrence
Cause
Action
• The pump cannot be turned on
when using AC power.
• The AC icon
is not
displayed even after the AC
power cable is connected.
• Check the connection of the AC power cable to the
pump and the power supply. (See page 29.)
• The AC power cable is
damaged.
• The AC power cable is
internally disconnected.
• Replace the AC power cable.
Battery Replacement Time/Failure alarm
Occurrence
Cause
Action
• The battery operation time is
short even after fully charged.
• The battery icon
is displayed.
• The battery icon does not
display 5 bars after charging
for 8 hours or more.
• The battery has deteriorated
or is damaged.
• Replace the internal battery.
•
and the battery charge
level appear alternately after
more than 8 hours of charging.
• The sub-battery has
deteriorated or is damaged.
• Replace the sub-battery.
Flow rate
Occurrence
Cause
Action
• Pressing the Start switch does
not start solution delivery.
• The infusion line is not open.
• The syringe is not installed
correctly.
1. Replace the syringe correctly.
(See pages 32 and 51.)
2. Prime the line, ensuring that the plunger is kept in the
slider to prevent the solution from being delivered to
the patient. (See pages 38 and 58.)
• An occlusion has occurred in
the infusion line.
• Eliminate the cause of occlusion. (See page 136.)
• Flow rate accuracy is low.
• Syringe has not been installed
correctly.
• Check the setting of the syringe.
(See pages 32 and 51.)
• The syringe in use does not
match the syringe brand set on
the pump.
• Check that the syringe brand setting matches the brand
of the syringe in use.
• The flow rate is not set
correctly.
• Correct the flow rate setting.
• There is large height
difference between the pump
and patient.
• Adjust the patient to the pump height differential.
Summary of Contents for TERUFUSION TE-SS830P
Page 163: ...163 Memo ...