49
Operation Procedure
Turning the Power Off
Power switch
3 seconds
or more
•
When the operation ends, press and hold the Power switch
for 3 seconds or more to turn the power off.
Notes
• While holding the Power switch, the power off buzzer sounds, then the LCD displays
count-down arrows and turns off.
• If the Power switch is pressed and held without ending the TCI mode, a buzzer
sounds indicating that the TCI mode is in use. The buzzer continues to sound for
approx. 3 seconds until the power is turned off.
Warning
• While in the TCI mode, never inject the drug to the same patient again after completion. Doing so may result in an
overdose.
Caution
• Do not turn off the power while the TCI mode is in use; otherwise, all important information on the current TCI mode, such
as the predicted concentration, will be erased.